Question
continuous, burst, dynamic, distortion, efficiency, thermal, clamp or protection
continuous, burst, dynamic, distortion, efficiency, thermal, clamp or protection
terminal voltage, current limit, wiring, sag, ripple, battery or bench source
resistance or impedance, hot value, inductance, cooling, energy and guarding
waveform, frequency, crest factor, ramp, duty cycle, channels and source state
topology, differential range, common mode, probe, filter and overload
shunt or probe ratio, bandwidth, phase, insertion, saturation and placement
distortion, compression, loss, temperature, limiting, protection and recovery
raw captures, calculations, uncertainty, report, version and regression trigger
reference generator or documented Wondom test project
→ verified line-level source and mute path
→ amplifier input
current-limited fused supply
→ terminal-voltage measurement
→ Kelvin supply shunt or AC/DC current probe
→ amplifier power terminals
amplifier output
→ guarded resistive load bank
→ differential voltage probe across load
→ output-current shunt or current probe
→ synchronized scope / analyzer channels
additional channels or logger
→ distortion endpoint
→ load-bank temperature
→ amplifier temperature
→ fan / protection / remote state
→ remote abort and electrical isolation
The enclosed two-input/four-output Wondom ADAU1701 can provide controlled routing, level, mute, slew, limiter experiments, and—when the loaded project includes one—a documented internal test generator. It is not the power analyzer, load bank, current sensor, differential probe, or emergency controller.
Use a verified differential probe, isolated instrument, or manufacturer-approved method. Start with the downstream amplifier disabled and validate the source and Wondom path at line level.
Choose one question: rated continuous output, short burst, dynamic output, clipping onset, THD+N endpoint, supply current, efficiency, thermal compression, protection, reactive-load real power, or another documented test. State whether the procedure follows ANSI/CTA-2006-D, IEC 60268-3, a modified method, or a project-specific method.
Record amplifier model, serial, hardware and firmware, input mode, channel configuration, bridge state, filters, bass boost, gain, limiter, fan mode, remote logic, mounting orientation and cooling. Photograph terminals and save the exact setup.
Record resistor construction, topology, tolerance, pulse rating, continuous rating, required mounting, cooling, terminal ratings and enclosure. Measure cold resistance, branch sharing and hot resistance. Define temperature and recovery limits before the first high-power run.
Identify single-ended, bridged, BTL, floating or another output topology. Set differential probe range, common-mode range, attenuation, bandwidth, filter, sample rate and record length. Zero, compensate and validate the probe before connecting to the energized test.
Choose the location and sensor. For a shunt, calculate:
Vshunt = I × Rshunt
Ploss = I² × Rshunt
Use Kelvin sense points, record hot resistance and verify insertion loss. For a current probe, zero or degauss, set range, verify polarity and deskew it against the voltage channel.
With a measured resistive load:
Pout = Vrms² / Rhot
Iout = Vrms / Rhot
Also capture current as a cross-check. Confirm the waveform, load temperature, supply condition, channel state and endpoint remain valid during the averaging interval.
Visible flattening is useful, but it is not the only endpoint. Record THD+N, harmonic spectrum, compression, limiter activity, source clipping, analyzer overload and protection. Use the endpoint defined by the test contract.
p(t) = v(t) × i(t)
Pavg = average of p(t)
S = Vrms × Irms
On a loudspeaker or reactive load, S is apparent power. Use synchronized and deskewed waveforms for average real power. Record impedance state, frequency, enclosure, level, temperature, excursion risk and measurement interval.
Pin = average of vsupply(t) × isupply(t)
η = Pout / Pin
Ploss = Pin - Pout
Match the input and output intervals. Include all amplifier supply paths and record fan, onboard DSP, lighting, remote and auxiliary-output states. Plot efficiency and loss versus output rather than preserving one class-based percentage.
Declare waveform, peak, RMS, crest factor, ramp, on-time, off-time, repetitions, starting temperature and total session duration. Track output compression, supply sag, efficiency, temperature, fan behavior, limiting and protection. Continue logging after output stops.
Do not rely on one fixed cooldown. Require the declared terminal voltage, load resistance, amplifier and load temperatures, compression, protection and supply conditions to return before another run.
Publish the result with load, supply, channels, signal, duration, endpoint, temperatures, instrument chain, bandwidth, filtering, calibration, number of runs, variation, uncertainty or limitation statement, raw evidence, first limiting mechanism and exact system revision.
| Finding | Do not assume | Measure next |
|---|---|---|
| scope shows a clean sine but power is low | amplifier is weak | terminal supply, load resistance, source headroom, probe scaling, channel mode |
| clamp watts exceed resistor calculation | amplifier creates extra power | current phase, sensor scaling, bandwidth, timing, hot resistance, apparent power |
| first burst is strong and later bursts fall | protection is defective | supply sag, storage depletion, temperature, compression, fan and duty cycle |
| efficiency calculates above 100 percent | special topology explains it | mismatched intervals, missing current path, apparent output power, probe ratio, phase |
| load bank survives one run | continuous rating is proven | hot resistance, element and terminal temperature, branch sharing, cumulative energy |
For the measured resistive load and waveform under the stated bandwidth, temperature and endpoint conditions.
It is apparent power unless the voltage-current phase and waveform correlation are included.
Input paths, output power type, sensor scaling, timing, phase or averaging intervals are wrong or incomplete.
The declared measured recovery criteria—not a universal waiting time.
The complete 12-question data set is stored beside this lesson for future interactive use.